excellent technical specifications have been a distinguishing mark of rohde & schwarz emi test receivers right from the beginning. each r&s®esib model continues this tradition with the following characteristics:
frequency range | |
input 1 | 20 hz to 7 ghz / 26.5 ghz / 40 ghz depending on model |
input 2 | 20 hz to 1 ghz |
preselection | 9 or 10 preselection filters depending on model can be switched off in analyzer mode |
resolution bandwidths (receiver and analyzer mode) | |
6 db bandwidths emi bandwidths |
10 hz to 10 mhz, in decimal steps 200 hz, 9 khz, 120 khz, 1 mhz acc. to cispr 16 tolerance |
resolution bandwidths (analyzer mode) | |
3 db bandwidths | 1 hz to 10 mhz, in steps 1/2/3/5 |
displayed average noise level | ||
receiver mode; 0 db rf attn.; preamp. 20 db |
r&s esib7 | r&s esib26 / 40 |
9 khz to 150 khz (rbw 200 hz) | –10 to –24 dbμv | –10 to –24 dbμv |
0.15 mhz to 30 mhz (rbw 9 khz) | –7 to –17 dbμv | –7 to –17 dbμv |
30 mhz to 1000 mhz (rbw 120 khz) | <–6 dbµv | <–3 dbµv |
1 ghz to 7 ghz (rbw 1 mhz) | <6 to 9 dbμv | <9 to 12 dbμv |
7 ghz to 26.5 ghz (rbw 1 mhz) |
–
|
<4 to 6 dbμv/6 to 9 dbμv |
26.5 ghz to 40 ghz (rbw 1 mhz) |
–
|
– /20 to 26 dbμv |
max. dynamic range (bw 1 hz) | ||
1 db compr. point/displayed noise floor | 162 db | 160 db |
total level uncertainty (preamp. off) | ||
analyzer mode | ≤ 1 ghz 1 ghz to 7 ghz |
≤ 0.5 db ≤ 1.0 db |
receiver mode | 150 khz to 1 ghz 1 ghz to 4.5 ghz 4.5 ghz to 7 ghz |
≤ 1.0 db ≤ 2.0 db ≤ 2.5 db |
measurement speed | |
sweep (analyzer mode) | 1 µs to 2.500 s (zero span) 5 ms to 1.000 s (span ≥ 10 hz) |
scan (receiver mode) | 100 μs to 1.000 s |
max. input level | |
rf attn. ≥ 10 db; sinewave ac voltage | 137 dbμv (= 1 w) |
max. input pulse voltage (10 µs) | |
input 2 (≤1 ghz); rf attn. ≥ 10 db | 1.500 v for the r&s esib7 250 v for the r&s esib26/40 |